• Home
  •  Product details
  • FAQ
  • Requirement
  • Contact Us

Nurse Call System ...

to ensure Patient Care

                                                                                                                                        www.datacareindia.com

Nurse Call System - OPD

Models of Nurse Call System:

 Elite+
 Enterprise
 Esteem-GSM
 Esteem-Sleek
 NCS-OPD


Description of NCS OPD

      The system is designed to help consulting Doctors working from their OPD rooms to call Nurse for any help. The system also allows for Doctors’ presence indication both on a common panel at Nurse station and on the Door indicator individually.
      The system consists of Dual Function Nurse Station Controller (NSC_DF), two types of displays (static LED windows and dynamic LCD display), CNP Module (DFRU), Toilet Calling Unit (CCTU) and a Door Indicator (DI)
  ncsopd_blk
For enhanced features like call forwarding, call escalation and centralized display following modules can be added to the system

  • Dual Function Supervisory Controller (SUC_DF)
  • Supervisory Display

Dual Door Indicator also serves two purposes namely: Doctor’s presence is indicated by lighting up the Door indicator in Green color while the absence is indicated in Red color. If a nurse is requested for, the Door indicator is made to flash in Amber color.

If LCD TV and management information is required then the PC has to be added to the system.The LCD TV can be made use of both for Nurse Call and Doctor’s Presence indication.

Optional Supervisory Display can work in the absence of PC for indicating call escalation.


Product Features:

  • Designed to carry out dual functionality for doctors operating from OPDs Viz: to seek assistance from Nurse Station and to indicate the presence of the doctor in that OPD.
  • Only one Cat5 cable is made use of to convey both information to Central Station.
  • Common door indicator used to show the presence of the doctor as well as a pending Call.
  • While for the doctors' presence indications, Static LED based windows are used. For indicating Origination of call, an LCD display windows is made use of.
  • The System can also provide for triggering call from the Toilets in the OPD Area.

NCS Specifications:

 

Modules:

NS-Display

OPD display               This is a Microcontroller based device placed in the nurse station with dynamic windows which displays the room number of the calling patient and also it has a voice module to announce the call. There are three buttons on it,

Power Button: Pressing this for long time(approximately 3 sec) turns the device ON or OFF.

MUTE Button: It is for acknowledging a call from patient. It will make the flashing window steady and also silence the voice module until a call is cleared.

Test Button: On pressing the button it will flash all the windows and will announce ” Attention, Patient calling ” to make sure that all the components are working fine.

Doctor Presence Indicating Panel (DPIP)

DPIP The static LED display is made use of to indicate doctor’s presence (up to 3 of 16 window panels) while the LCD display is made use of for Nurse Call indication.

Technical specifications:
Dimension :
Weight : 3 Kg
Material : Plastic Enclosure

CNP Module (DFRU)

                     CNP Module (DFRU) is designed around a RISC based microcontroller to read the keys on it, to interface with all room modules viz: TCU, DI and to report their status to the Nurse Station Controller.
                                           Features:
DFRU
  • Houses 5 feather touch keys viz: CALL, ACK, CLEAR for closing the call, DOCTOR-IN & DOCTOR-OUT.
  • Gets integrated with the Palm switch for initiating a call easily
  • Makes use of one multicolor LED for indicating Call Status and two more LEDs for Call Acknowledgement and Doctor Presence indication.

Specifications:
      Gets powered by Nurse Station Controller, consumes less than 5 watts. Uses 4 pairs of wires to get inter-connected with Nurse Station controller.

Technical specifications:
Dimension : 118mm (L) X 90mm (W) X 53mm (H)
Weight : 200 g
Material : Plastic Enclosure

Toilet Calling Unit (CCTU)

CCTU
  • CCTU houses a pull cord switch to trigger a Call.
  • The unit comes with 2 feather touch keys viz: CARE and CLEAR.
  • It has an integrated dual color LED indicator to show Call status.

Technical specifications:
Dimension :118mm (L) X 90mm (W) X 53mm (H)
Weight : 115 g
Material : Plastic Enclosure

Dual Function Nurse Station Controller (NSC_DF)

Controller             Designed around a high speed RISC based microcontroller to interface with 16 room units and four nurse station(NS) display modules. It collects information from all the enabled room units, sends indicator information to all the calls and accordingly instructs Nurse station(NS) display modules. It also sends status to the Supervisor controller about all the activities.

Dual Function Supervisory Controller (SUC_DF)

Supervisory Controller             Designed making use of a high speed RISC based micro controller to get connected to 16 nurse station controllers and to a PC. It interrogates all the nurse Controllers to get information related to all the activities of different room units and process them to arrive at Call forwarding, Call escalation information which are forwarded to a PC, to a common large size display unit and to the Nurse station displays through Nurse station controller. It also reports to the PC about all the other activities.

Supervisory LED Display

Supervisory Display             This Display system is again designed around a microcontroller to get connected to the supervisory controller to receive any information to be shown and to put such information on the LED matrix. The display size is designed to be able to read it from a distance of 100 feet easily. The supervisory controller uses this display to show call escalated, call forwarded etc.,


Door Indicator

            These semi circular shaped indicators can be mounted on the wall near the patient's door so that , nurse walking through the corridor can see it from a distance.These are made of super bright LEDs to indicate patient's call pending which can be seen from the corridor itself. Door indicator

Specifications:
Dimension :100mm (diameter) X 60mm (H)
Weight :73 g
Material :Plastic Enclosure



Models of Nurse Call System:

Download PDF  Elite+
 Enterprise
 Esteem-GSM
 Esteem-Sleek
 NCS-OPD


Nurse Station Display:

Indicators         : flashing 5 mm LED to indicate NCS is active.
Power                : 230 V AC +/-10%,50Hz, 15 watts

Display Windows
  • LCD display                   : 320x240 pixels
  • Dimension                     : 135W mm X 110H mm
  • Symbol Size                  : 64x64 pixels on NS Display.

Front Panel Keys
  • ”MUTE” for silencing the voice
  • ”TEST” for testing all the windows and the voice module
  • ”PWR” to turn ON/OFF the equipment

Dimension/Weight:
  • NS-display                         : 255W X 205H X 110D 2Kg (approx.)
  • Calling Unit                       : 118mm (L) X 90mm (W) X 53mm (H) 160 gm (approx.)
  • Door Indicator                  : 100mm (diameter) X 60mm (H) 73 gm (approx.)

Cable Specifications:
  • Standard                             : Belden #8795 equivalent
  • Thickness                           : 22AWG twisted pair Characteristic
  • Impedance (Zo)                 : 100 E
  • Distributed Capacitance : 15 pF/ft
  • Propagation Delay            : 1.3 nsec/ft
  • Conductor Resistance    : 0.027 Ohms/ft (90 Ohms/Km approx.)

Environmental Specifications for all Modules:
  • Operating Temperature  : 0 to 50* C
  • Humidity                              : 95% at 40*C
  • Storage Temperature      : 0 to 70* C
  • Altitude                                : 10,000 ft

Copy right © 2018  DATA CARE GROUP